metal artifact reduction sequence

A metal artifact reduction sequence (MARS) is intended to reduce the size and intensity of susceptibility artifacts resulting from magnetic field distortion.

A variety of techniques are used for reducing metal artifacts at MRI, both for addressing artifacts due to the presence of metal in the image plane (in-plane artifacts) and for artifacts due to metal in an adjacent plane (through-plane artifacts).

In-plane artifact reduction

A number of simple changes to the scan protocol can greatly reduce artifacts. Examples are :

  • lower magnetic field strength: 1.5 T rather than 3 T
  • increase bandwidth during slice selection and readout
  • increase matrix: 512 pixel
  • maintain good signal to noise ratio by increasing number of excitations (NEX)
  • spin echo instead of gradient echo where possible
  • STIR for fat suppression (spectral frequency selective fat suppression performs better in a homogeneous field)
  • shorter echo spacing
  • smaller water-fat shift
  • thinner slices
  • view-angle-tilting (VAT)
Through-plane artifact reduction

Multiacquisition variable-resonance image combination (MAVRIC) is a specialized sequence designed by GE to minimize metallic artifact around metallic prostheses . It relies on 3D FSE sequences, using multiple different overlapping volumes at different frequency offsets.

Another technique developed by Siemens used for addressing through-plane metal artifacts is Slice-Encoding for Metal Artifact Correction (SEMAC), where an additional slice-encoding gradient is added to a standard fast-spin echo sequence . The combination of the MAVRIC and SEMAC technique is known as MAVRIC-SL .